S1500
The ventilator makes a good performance in operation room,ICU department and emergency treatment.
Application
The ventilator makes a good performance in operation room,ICU department and emergency treatment.
It used to assist or replace the spontaneous breathing for adult pediatric and neonatal more than 2kg.
25 years experience in Market-oriented ventilator make us professional and reliable, satisfying all your needs in ventilation.
Due to the flexible configuration, good quality and competitive price, S1100 has soon become the superstar of market.
Features
- 12.1” TFT screen displays the ventilation parameters, alarm information, and oscillograms, make every operation more easily.
- Multiple ventilation mode can meet different clinical requirements.( IPPV, A/C, PCV, SIMV,PSV, SPONT/CPAP,SIGH,MANUAL) 3 oscillograms for your choice, 2 of them can be displayed on the screen at the same time.
- Humidifier can heat and wet breathing gas, makes it comfortable for patient to breathe.
- Rapid oxygen supply,automatically offer high flow rate oxygen within two minutes.
- High temperature resistance breathing circuit is reusable and anti-pollution.
Safety
- 14 types of sound and visual alarm information, easier for users to do some error checking and troubleshooting.
- Built-in oxygen concentration sensor, ensure stable precision of oxygen concentration.
- Easy to move with four casters, easy to stop with two brakes.
- Separate design of electronic circuit and gas flow rate keep safe running of ventilator.
- Compact long life internal battery can provide emergency power,avoid risk of patient.
- Self-check before operation,eliminate system mistake.
TECHNICAL SPECIFICATION | Value | TECHNICAL SPECIFICATION | Value |
---|---|---|---|
Ventilation mode | Alarm and protection | ||
IPPV, A/C, PCV, SIMV, SPONT/CPAP, PSV, SIGH, MANU | The AC power failure alarm | Power failure or no connection | |
Internal backup batery low voltage alarm | ≤ 11.3 ± 0.3 V | ||
Ventilator parameter | No tidal volume | No tidal volume within 6 s | |
Tidal volume (Vt) | 0, 20 ~ 2000 mL | High minute volume alarm | 5 L/min ~ 99 L/min |
Frequency (Freq) | 1 min ~ 100 min | Low minute volume alarm | 1 L/min ~ 30 L/min |
Oxygen concentration | 21 % ~ 100 % | High airway pressure alarm | 20 cmH2O ~ 100 cmH2O |
I:E | 4 : 1 ~ 1 : 8 | Low airway pressure alarm | 0 cmH2O ~ 20 cmH2O |
PEEP | 0 cmH2O ~ 40 cmH2O | High oxygen concentration alarm | 19 % ~ 100 % |
Pressure limit | 20 cmH2O ~ 100 cmH2O | Low oxygen concentration alarm | 18 % ~ 99 % |
Pressure triggering sensitivity(Ptr) | -20 cmH2O ~ 0 cmH2O(Based on PEEP) | Continuous pressure alarm | (PEEP + 1.5 cmH2O) over 16s |
Flow triggering sensitivity(Ftr) | 0.5 cmH2O ~ 30 cmH2O | Suffocation warning | 5 ~ 60 s |
SIGH | 0 (off) 1/100 ~ 5/100 | Fan error | Show on screen |
Apnea ventilation | OFF, 5 s ~ 60 s | Oxygen deficit | Show on screen |
The maximum limited pressure | < 12.5 kPa | ||
Monitoring parameter | |||
Frequency (Freq) | 0/min ~ 100/min | Working condition | |
Tidal volume (Vt) | 0 mL ~ 2500 mL | Gas source | O2, Air |
MV | 0 L/min ~ 99 L/min | Pressure | 280 kPa - 600 kPa |
Airway pressure | 0 cmH2O ~ 100 cmH2O | Voltage | 220 V ± 22V |
Dynamic lung compliance monitoring | 1 mL/cmH2O ~ 1000 mL/cmH2O | Power frequency | 50 Hz ± 1Hz |
Oxygen concentration | 15 % ~ 100 % | Input power | 900 VA(With air compressor) |
250 VA(Without air compressor) | |||
Packing size | |||
Main engine: L 560 * W 560 * H 605 mm | Oscillogram | ||
G.W. : 35 KG, N.W. : 17 KG | P-T(Pressure-Time) | ||
Air compressor: L 683* W 687 * H 1140 mm | F-T(Flow-Time) | ||
G.W. : 100 KG, N.W. : 65 KG | P-V Loop(Pressure-Volume Loop) |




To better ensure the safety of goods, professional, environmentally friendly, convenient and efficient packaging services will be provided.
Wooden case packing size:
Main components: L 560 * W 560 * H 605 mm
G.W. : 40 KG, N.W. : 20 KG
Air compressor: L 670* W 700 * H 1160 mm
G.W. : 100 KG, N.W. : 64 KG
Reviews
There are no reviews yet.